CardioTouch 2000
CardioTouch 2000
The CardioCare 2000 is an economically priced interpretive 12 channel electrocardiogram (ECG/EKG) machine that does not compromise on performance. Suitable for use in private practice, the ER, or hospitals. The CardioCare 2000 is designed for maximum ease of use and convenience. Software upgradeable and networkable, the CardioCare 2000 is an excellent choice for your diagnostic needs.
Simultaneous 12 Channel Electrocardiograph (ECG/EKG)
The CardioCare 2000 acquires 10 full seconds of ECG data simultaneously from 10 leads to produce an accurate 12 channel printout with analysis.
ECG Interpretation Program
Over 80 kinds of interpretation results based on the advanced Minnesota Code are provided to aid with your diagnosis.
The interpretation algorithms are expanded with regular software updates.
A4 Full Size Paper
The CardioCare 2000 will produce full size A4 printouts in 1, 3, 6, and 12 channel formats.
Full size printouts improve readability and make it easier for filing in patient charts.
A convenient Copy button lets you generate ECG copies in any of the available formats.
| Feature | Specification |
|---|---|
| ECG Leads | 10 lead 12 channel simulataneous 10-second resting ECG |
| Dimension | 296 (w) x 305.5 (h) x 92.5 (d) mm |
| Approximately 2.98 kg | |
| Recording Channel | 3, 6, 12 channel and 1 channel (60 sec) |
| Sensitivity | 5, 10, 20 auto (I~aVF: 10, V1~V6:5) mm/mV |
| Printing Speed | 12.5, 25, 50 mm/s |
| Filters | AC (50/60 Hz, -20dB or better) |
| Muscle (25~35Hz, -3dB or better) | |
| Base Line Drift (0.1Hz, -3dB or better) | |
| Low pass filter (off, 40Hz, 100Hz, 150Hz) | |
| Display | 2 x 16 character LCD display |
| LCD Display | Sensitivity, speed, filter status, HR, printing form, rhythm lead |
| LED Indicator | Signal quality, power source, battery |
| Keyboard | Membrane keyboard |
| Alphanumeric and symbol available | |
| Patient Data | ID, name, age, sex, height, weight |
| Basic Measurement | Heart rate, PR, QRS, QT/QTc, P-R-T axis |
| Recorder | Thermal print head, thermal roll paper |
| Report papers: | |
| -Width: A4: 210mm or 8.5″ | |
| -Length: A4: 300mm or 11″ | |
| Effective writing width: 204mm | |
| Resolution: | |
| – Vertical: 8 dots/mm | |
| – Horizontal: 16 dots/mm | |
| Electrical | Internal noise: 20uV(p-p) max |
| Input circuit: Floating input | |
| Isolated and defibrillation protected | |
| Input impedance: >10MQ | |
| Input range: > +/- 5mV | |
| Common Mode Rejection: > +100dB | |
| DC offset voltage: > +/- 300mV | |
| Time constant: 3.2 sec | |
| Patient leakage current: | |
| Frequency response: 0.05~150Hz | |
| Signal Quality Control | Disconnected lead detection |
| Power | Power Supply: AC or built-in Battery (option) |
| Power requirement: 100 – 240 Vac, 50/60Hz, 1.0 – 0.5A | |
| Power consumption: 60w max | |
| Battery Capacity | 1 hour of normal use (approximately 100 automatic ECG printouts) |
| Communication | PC connection with RS-232 interface and LAN |
| Safety | Class 1, Type BF |
| Conformity | CE, CSA |
| Environment | Operating humidity: 30~85% |
| Operating temperature: 10° ~40° C | |
| Atmospheric pressure: 70 ~ 106 KPa | |
| Standard Accessories | Power Cable – 1 ea Patient Cable – 1 ea Electrode – 10 ea Chart Paper – 1 ea Operation Manual – 1 ea |
| Options | Rechargeable Battery (ni-mh), Cart, Hanger, Bag, PC s/w, Limb electrode, Chest electrode, RS-232C null cagle, LAN cable (Ethernet-RJ-45) |